RP-HPLC determination of lycopene in microcapsules

Ding-Qiang Lu*, Xiao-Yun Lu, Kai-Feng Lu, Ji-Jun Hu   

  1. College of Life Science and Pharmaceutical Engineering, Nanjing University of Technology, Nanjing 210009, China
  • Received:2006-09-05 Revised:2007-02-10 Online:2007-03-15 Published:2007-03-15
  • Contact: Ding-Qiang Lu*

Abstract: Aim A RP-HPLC method for determination of lycopene in microcapsules was established. Methods The HPLC assay was performed on an Alltima C18(4.6 mm × 250 mm, 5 μm) column with a mixture of methanol-THF-water (66:30:4, V/V/V) as mobile phase at a flow rate of 1.5 mL·min-1 and the UV detection wavelength was 472 nm. Results The linear range of lycopene was 3.6-18 μg·mL-1, r = 0.9998, the average recovery was from 99.81% to 101.06% with RSD less than 1.83%. The RSD of intra-day and inter-day precision were less than 3.34%. Conclusion The method is simple, accurate and suitable for the determination of lycopene in microcapsules.

Key words: RP-HPLC, RP-HPLC, Lycopene, Lycopene, Microcapsules, Microcapsules